Abstract
Objective
The growing number of technical devices in ICUs makes noise exposure a major stressor. The purpose of this study was to assess noise levels during routine operation in our ICU.
Design
Our ICU is an open ward with four rooms, constructed in the 1960s. During the study period, 4 patients were in the controlled room and were treated by 4 nurses during the day and by 2 at night. A-weighted sound pressure levels (SPL) were measured continuously for 2 days and nights. Also measured were the alarms of various appliances. For gross overall evaluation it is customary to state the Leq, i.e. the energy-averaged level during measurement. The annoyance caused by noise depends more on rare events of high intensity. Therefore, the distribution of SPL values (Ln) over time was also analysed.
Results
SPL was roughly the same during the day and at night, with Leq between 60–65 dB(A) and peaks up to 96 dB(A). Most alarms reach an SPL of 60–70 dB(A), but some exceed 80 dB(A). During teaching rounds Leq exceeds 65 dB(A).
Conclusion
During the day and at night SPL always surpasses the permissible noise exposure for 24 h of 45 db(A) recommended by the US Environmental Protection Agency. Alarms cause the most irritating noise. Hospital management should pay attention to internal noise, and SPL should be measured routinely.
Article PDF
References
Aitken RJ (1982) Quantitative noise analysis in a modern hospital. Arch Environ Health 37:361–364
Sharkey T (1988) Sounding an alarm on alarms (letter). Am J Nurs 88:33
Hansell HN (1984) The behavioral effects of noise on man: the patient with “intensive care unit psychosis”. Heart Lung 13:59–63
Falk SA, Woods NF (1973) Hospital noise levels and potential health hazards. N Engl J Med 289:774–781
US Environmental Protection Agency (1974) Information on levels of environmental noise requisite to protect health and welfare with an adequate margin of safety. EPA, Washington DC
Field T (1990) Alleviating stress in newborn infants in the intensive care unit. Clin Perinatal 17:1–5
Thomas KA (1989) How the NICU environment sounds to a preborn infant. MCN 14:249–252
Hodge B, Thompson JF (1990) Noise pollution in the operating theatre. Lancet 335:891–894
Hilton A (1987) The hospital racket: how noisy is your unit? Ann Y Nurs 1:59–60
Kerr JH, Hayes B (1983) An “alarming” situation in the intensive therapy unit. Intensive Care Med 9:103–104
Shapiro RA, Berland T (1972) Noise in the operating room. N Engl J Med 287:1236–1238
Updike P (1990) Music therapy results for ICU patients. Dimens Crit Care Nurs 9:39–42
Pownall M (1987) Medical alarms: ringing the changes. Nurs Times 83:18–20