Crisis management during anaesthesia: vascular access problems
Free
  1. R J Singleton1,
  2. S B Kinnear2,
  3. M Currie3,
  4. S C Helps4
  1. 1Senior Staff Specialist, Department of Anaesthesia and Intensive Care, Royal Adelaide Hospital and University of Adelaide, Adelaide, South Australia, Australia
  2. 2Consultant Anaesthetist, private practice, Adelaide, South Australia; Visiting Anaesthetist, Flinders Medical Centre, Bedford Park, South Australia, Australia
  3. 3Clinical Quality Coordinator, Goulburn Base Hospital, Goulburn; and Consultant in Clinical Quality, Southern Area Health Service, New South Wales, Australia
  4. 4Metabolic Neurochemistry Unit, Department of Medical Biochemistry, School of Medicine, Flinders University, Bedford Park, South Australia, Australia
  1. Correspondence to:
 Professor W B Runciman
 President, Australian Patient Safety Foundation, GPO Box 400, Adelaide, South Australia, 5001, Australia; researchapsf.net.au

Abstract

Background: In confronting an evolving crisis, the anaesthetist should consider the vascular catheter as a potential cause, abandoning assumptions that the device has been satisfactorily placed and is functioning correctly.

Objectives: To examine the role of a previously described core algorithm “COVER ABCD–A SWIFT CHECK”, supplemented by a specific sub-algorithm for vascular access problems, in the management of crises occurring in association with anaesthesia.

Methods: The potential performance of a structured approach was evaluated for each of the relevant incidents among the first 4000 reported to the Australian Incident Monitoring Study (AIMS).

Results: There were 128 incidents involving problems related to vascular access. The structured approach begins distally, checking the infusion device or fluid (12 incidents), moving proximally by way of the fluid giving line (10), the line deadspace (8), then the catheter/skin interface (65), and on to the peripheral vascular tree (3) and central venous space (23), and finally, the interface of the vascular access system and the attending staff (7). The approach was able to accommodate all the vascular access problems among the first 4000 incidents reported to AIMS.

Conclusion: The approach has potential as an easily remembered and applied clinical tool to lead to early resolution of vascular access problems occurring during anaesthesia.

  • vascular access
  • crisis management
  • line deadspace
  • equipment problems
  • anaesthesia complications

Statistics from Altmetric.com

Request Permissions

If you wish to reuse any or all of this article please use the link below which will take you to the Copyright Clearance Center’s RightsLink service. You will be able to get a quick price and instant permission to reuse the content in many different ways.

The cause of a perioperative crisis is not always readily apparent to the anaesthetist. A problem may arise directly from the immediate actions of the attending theatre staff, or it may result from a later complication of a previous intervention. In this latter event, because the crisis may have arisen from actions taken before the patient reached the operating room, the potential exists for the cause of the problem to be beyond the anaesthetist’s immediate frame of reference, further hindering the diagnostic process.

The great majority of patients presenting for anaesthesia and surgery require the establishment and ongoing use of some type of vascular access. Patients may present to anaesthetists with vascular access already apparently established, or the anaesthetist may be required to establish the access. In confronting an evolving crisis situation, the anaesthetist should always consider the ubiquitous vascular catheter as a potential cause for the emergency, abandoning any assumption that the device has been satisfactorily placed and is functioning correctly.

In 1993, a “core” crisis management algorithm, represented by the mnemonic COVER ABCD–A SWIFT CHECK (the AB precedes COVER for the non-intubated patient), was proposed as the basis for a systematic approach to any crisis during anaesthesia where it is not immediately obvious what should be done, or where actions taken have failed to remedy the situation.1 This was validated against the first 2000 incidents reported to the Australian Incident Monitoring Study (AIMS). AIMS is an ongoing study which involves the voluntary, anonymous reporting of any unintended incident which reduced, or could have reduced, the safety margin for a patient.2

It was concluded that if this algorithm had been correctly applied, a functional diagnosis would have been reached in 40–60 seconds in 99% of applicable incidents, and the learned sequence of actions recommended by the COVER portion would have led to appropriate steps being taken to handle the 60% of problems relevant to this portion of the algorithm.1 However, this study also showed that the 40% of problems represented by the remainder of the algorithm, ABCD–A SWIFT CHECK, were not always promptly diagnosed or appropriately managed.1–3 It was decided that it would be useful, for these remaining problems, to develop a set of sub-algorithms in an easy to use crisis management manual.4 This study reports on the place of the COVER ABCD–A SWIFT CHECK algorithm in the diagnosis and initial management of vascular access problems and provides an outline of a specific crisis management sub-algorithm for these problems during anaesthesia, based on a geographical approach to the patient, and provides an indication of the potential value of using this structured approach.

METHODS

Of the first 4000 incidents reported to AIMS, those that made reference to incidents involving vascular access were extracted and analysed for relevance, presenting features, type of surgery, cause, management, and outcome. The COVER ABCD–A SWIFT CHECK algorithm, described elsewhere in this set of articles,4 was applied to each relevant report to determine the stages at which the problem might have been diagnosed and to confirm that activating the COVER portion would have led to appropriate initial steps being taken. As vascular access problems are not dealt with by this algorithm, a specific sub-algorithm was developed for these problems (see figure), and its putative effectiveness was tested against the reports. The potential value of this structured approach (that is, the application of COVER ABCD–A SWIFT CHECK to the diagnosis and initial management of this problem, and the application of the relevant sub-algorithm for vascular access problems) was assessed in the light of AIMS reports by comparing its potential effectiveness for each incident with that of the actual management, as recorded in each report.

RESULTS

Of the first 4000 reports reported to AIMS, 128 described incidents involving vascular access. Applying the mnemonic of the crisis management algorithm, “COVER ABCD”, to these incidents, all belonged to the category covered by “R”—“review all (vascular) equipment”. As this is a very broad category encompassing many problems, these incidents were further analysed using a sub-algorithm based on a “geographical” approach to the patient (fig 1). Using this sub-algorithm the anaesthetist proceeds with a systematic review of vascular access equipment beginning distally at the infusion device or fluid, and moving proximally by the way of the fluid giving line, the line deadspace, then the catheter/skin interface, and onto the peripheral and central vascular tree. The incidents reported to AIMS are presented in groups below, according to categories based upon the use of the described sub-algorithm.

Figure 1

 Vascular access problems: a sub-algorithm.

1. Infusion device or fluid (12 cases)

In four cases, a problem arose because the wrong infusion pump or solution had been inadvertently connected to the patient. In a further three reports, incidents occurred because an infusion pump was giving an unrecognised overdose. In three cases, anaphylaxis to the infusion fluid was finally recognised as the cause of the problem. In two patients, a dose of a drug from the deadspace of an infusion pump was eventually recognised as the causative event of an incident.

2. Fluid line and equipment problems (10 cases)

Five arterial lines and three central venous lines were faulty in either their manufacture or assembly at the bedside, thus creating problems in their insertion or use for the measurement of vascular pressures. Similar problems occurred with two peripheral lines: in one case, poor preparation almost allowed a large air embolus from an unprimed blood warming coil, and in another case a leak at the hub of an intravenous cannula put the patient at increased risk.

3. Line deadspace (eight cases)

In six patients, drug remaining in the deadspace of the intravenous cannula caused late muscle paralysis upon subsequent injection of fluid through the vascular port. Succinylcholine was the drug in all cases.

There were two incident reports describing anaesthetists inadvertently mixing two drug solutions by their sequential injection into the deadspace of the intravenous line, resulting in precipitation in the line deadspace, causing occlusion at an inopportune time. On both occasions, the drugs were thiopentone and vecuronium.

4. Catheter/skin interface (65 cases)

There were 21 incidents of intravenous catheters assumed to be intravascular which proved to be located extravascularly upon attempted use. In 10 of these, the anaesthetist did not insert the catheter himself. There were 19 cases of assumed intravenous catheter placement actually later realised to be intra-arterial in position. Twelve incident reports described extravasation or dislodgment of a catheter upon unplanned movement of the patient. These catheters were not adequately secured. There were eight cases of a vascular catheter becoming unknowingly disconnected causing unrecognised blood loss; these incidents occurred when the catheter was hidden from the anaesthetist by the surgical drapes. Two incident reports described fragments of catheter remaining in the patient upon device removal. There was a single incident of an intravenous catheter being placed inappropriately on an arm with an arterio-venous fistula, another of the catheter insertion being delegated to an inappropriately trained individual, and one incident of continued inadvertent intra-arterial injection of drug despite considerable pain on injection.

5. Peripheral vascular tree (three cases)

There was one incident report of an arterial line causing thrombosis, and one report of a radial arterial line producing hand ischaemia. Another incident involved an intravenous catheter in the antecubital fossa damaging the brachial artery.

6. Central venous space (23 cases)

There were five reports arising from attempted central venous catheterisation via an internal jugular vein. Three incidents described puncture of the cuff of the endotracheal tube by the advancing needle of an attempted jugular cannulation. Two other incidents described the development of a large neck haematoma.

Central venous cannulation by the subclavian route produced five reports of pneumothorax, four reports of a pleural effusion or hydrothorax, and one case of haemothorax. There was one central venous and one central arterial thrombosis reported, and one description of cardiac tamponade after subclavian needle passage. One incident outlined a failed subclavian vein catheterisation and the reporter attributed the lack of success to unrecognised hypovolaemia. Four incidents described the acute onset of dysrhythmia with the insertion of a central venous catheter.

7. Staff and situation problems (seven cases)

Three incidents described the loss of function of the vascular catheter produced by the surgeon leaning on the equipment site. Three needlestick injuries were reported, and in each instance the injury was suffered not by the clinician inserting the needle, but by an assistant. One incident arose because of the failure of the staff at the transfer of the patient to communicate the presence of an intra-arterial catheter.

DISCUSSION

This descriptive paper illustrates that the problems associated with vascular access are many and varied in nature, and may not necessarily arise solely from the actions of the anaesthetist. In any crisis or unsolved patient dilemma, the clinician should always consider the vascular access system as a potential cause for the problem. In the first instance, this will be accomplished by applying the crisis management mnemonic “COVER ABCD”. In reviewing all vascular access equipment, the anaesthetist requires a systematic approach to reduce the likelihood of the problem remaining undetected, and the sub-algorithm described above is offered as a means of proceeding in this manner. This geographical approach can be commended on the strength of its being able to accommodate all the vascular access problems reported to AIMS among the first 4000 incidents. It is hoped that it will also prove to be an easily remembered and applied clinical tool, even during times of acute occupational stress for the anaesthetist.

Key messages

  • Of the first 4000 anaesthesia incident reports to AIMS there were 128 incidents involving problems related to vascular access.

  • All of these incidents belonged to the “R”—“review all (vascular) equipment” step in the COVER ABCD algorithm. The further application of the specific vascular access sub-algorithm is based upon a “geographical” approach with the anaesthetist’s checking beginning distally and moving centrally stepwise.

  • The resultant categories were: infusion device or fluid (12 cases); fluid line and equipment problems (10 cases); line deadspace (8 cases); catheter/skin interface (65 cases); peripheral vascular tree (3 cases); central venous space (23 cases); staff and situation problems (7 cases).

  • The resolution of some vascular access problems may be facilitated by the additional application of the sub-algorithm for problems related to drug administration, described elsewhere in the series of papers.

  • Twelve reports of extravasation or dislodgement of a catheter were the result of unplanned patient movement.

  • Subclavian venous cannulation produced five reports of pneumothorax, four reports of pleural effusion or hydrothorax, and one case of haemothorax.

  • Three incidents of loss of catheter function resulted from the surgeon leaning on the site.

  • The correct application of the core algorithm, followed by the specific sub-algorithm employing the geographical approach, was considered to accommodate all the vascular access problems among the first 4000 AIMS incidents.

As a complementary aide memoire in connection with incidents related to vascular access linked to drug administration problems in anaesthesia, the relevant pages developed for the Crisis Management Manual5 are displayed in the figure. These are designed to be consulted for the first time during private study time. However in the resolution of a difficult vascular access related problem, when assistance is available, they may play a role along with the other sub-algorithms in resolving an unusual, obscure, or difficult situation.

Finally, it is important that a full explanation of what happened be given to the patient, and that the problem be clearly documented in the anaesthetic record. If a particular precipitating event was significant, or a particular action was useful in resolving the crisis, this should be clearly explained and documented.

Acknowledgments

The authors would like to thank all the anaesthetists in Australia and New Zealand who contributed to the 4000 incident reports upon which this and the other 24 papers in the Crisis Management Series are based. The coordinators of the project also thank Liz Brown for preparing the draft of the original Crisis Management Manual; Loretta Smyth for typing; Monika Bullock, RN, for earlier coding and classifying of data; Dr Charles Bradfield for the electronic version of the algorithms; Dr Klee Benveniste for literature research; Drs Klee Benveniste, Michal Kluger, John Williamson, and Andrew Paix for editing and checking manuscripts.